Liquid
Chromatography
Click here to subscribe to the EAC's LC-MS System listserv.
Hewlett-Packard
(Agilent) 1100 LC-MSD
Our
high performance liquid chromatography system is equipped with
an in-line solvent degasser, two-channel gradient pump, 96 sample
auto-injector, 2 column oven, UV-Vis DAD, and quadrupole mass
spectrometer. We have both the Electrospray Ionization (ESI)
and the Atmospheric Pressure Chemical Ionization (APCI) interfaces.
Instrument
Reservations
You
must note times you are using on the web
calendar. This is so that users from across campus won't
have to walk all the way to the lab to find out it is in use.
Loop
Injection of Samples
If
you do not need to chromatographically resolve the components
of your sample, or you have a nearly pure sample you can bypass
the column to get an ESI-MS or APCI-MS spectrum of your sample
directly.
Reverse
Phase Chromatography
We
have various LC columns of various lengths, diameters, particle
size, and phase available for public use. Most of these columns
can handle solvent conditions from 20% to 100% organic. You
should consult the manufacturers' websites or the published
chemical literature for assistance in determining what column
phase will effect the separation you desire.
When
selecting buffers for use with ESI-MS or APCI-MS be sure to
choose volatile buffers. Non-volatile buffers like sodium phosphate
will deposit salt inside the ion source, reducing the efficiency
of ion production.
The
EAC provides basic solvents including methanol, acetonitrile,
and ultra-pure water. You must provide other solvents or buffers.
Diode-Array
Detection UV-VIS Spectroscopy
The
UV-VIS detector is equipped with a high-pressure, high-sensitivity
cell from the manufacturer. Because most users are not actually
using the the UV-VIS data, the detector is normally kept disconnected
from the path of solvent flow. If you need the UV-VIS data,
contact EAC staff for training in connecting and disconnecting
the UV-VIS.
Mass
Spectrometer
The
mass spectrometer is a quadrupole design with a mass range to
2,000 Daltons. No MS-MS is possible on this instrument, however
the ion kinetic energy can be varied through the interface region
allowing for greater or lesser degrees of fragmentation.